54
Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
7.1.3.8.3 DPG II Models 14000/15000 Terse Message Set Emulation
The following device dependent messages include the original terms enabled for the
DPGII. Most of these terms have an equivalent message in the above, expanded mes-
sage list. When sending a terse message to the CPG 2500, transmit the message fol-
lowed by an X. The X signals the CPG 2500 to execute the command contained in the
message.
A Command
(used with RATE option; same as RATE_WINDOW command):
The A command sets the pressure rate filter window from 0 to 9 percent of full scale.
The command format is: AnX where n = 0 to 9
C Command
no equivalent expanded command): The C command is not emulated.
E Command
(no equivalent expanded command): The E command is not emulated.
F Command
(used with RATE option; same as RATE_FILTER command)
Q Command
(similar to some expanded commands):
The Q command is a request for data from the CPG 2500. The output data will be for-
matted according to the specific form of the Q command. Output formats 0 and 1 will
remain selected until changed by a subsequent Q command. Output formats 2 through
7 will be in effect for one output cycle only, after which the format will revert to 0 or 1,
whichever was last being used.
The syntax for the Q command is QnX where n is a number (0 - 7) as described in the
following table. The table also lists the resulting output format for each value of n.
space = an ASCII space character (32 dec)
<cr> = an ASCII carriage return (13 dec)
<lf> = an ASCII linefeed (10 dec)
EOI (End Of Instruction) is set with the <lf> (on the GPIB)
i
Notice
The Q1X command changes the output units to counts, and Q0X
resets it to default units.