Mensor/WIKA Operating Instructions Series CPG 2500
51
Digital Pressure Gauge
SERIES CPG 2500
SPAN
SPAN sp <value> lf
Use the SPAN command to correct the pressure
reading at FS. This command first requires the
PW password to be issued. Send the true pres-
sure value while maintaining at least 50% of the
FS pressure on the pressure port. (Refer to the
Maintenance and Calibration sections for details
on calibrating an instrument.)
Where <value> is true pressure value in current
engineering units.
SPAN?
SPAN? lf
The SPAN query will return the span correction
scale factor. Returns <value><cr><lf> where
<value> is a multiplication factor from 0.90000 to
1.10000.
TARE
TARE sp <value> lf
TARE sets the tare offset in the current engineering
units. Either the PWT or the PW password must
be sent before the first occurrence of the TARE
command. Where <value> is a value b/-
7.0000 psi.
TARE?
TARE?lf
Returns the current tare calibration variable.
Returns <value><cr><lf> where <value> is the
current TARE offset in the current units.
TYPE?
type?lf
Returns the type of pressure sensor in the instru-
ment:
ABSOLUTE PRESSURE<cr><lf> for an absolute
sensor,
or: GAUGE PRESSURE<cr><lf> for a gauge pres-
sure sensor.
UNITS
UNITS sp<value> lf
This command selects the engineering units to be
output on the bus and the display for all subse-
quent pressure readings.
Where <value> is a one or two digit units code
number from the ‘n’ column of Table 7.1.3.6.
UNITS?
units?lf
This query returns the units code and the ASCII
string for the units as <value><text><cr><lf>
where <value> is: the units code from column “n”
(see Table 7.1.3.6) and where <text> is the corre-
sponding ascii string listed under ‘Output Format.