Mensor/WIKA Operating Instructions Series CPG 2500
31
Digital Pressure Gauge
SERIES CPG 2500
6.5.2.3 Adjust Tab
This tab is for adjusting zero and head pressure correction and setting access to those
features. These features can be locked out by pressing the key icon. A password screen
will appear. Enter the correct password. Next select the utility tab. The padlocks can
then be toggled on or off.
6.5.2.3.1 Zero
Zero can be adjusted on this tab and the main screen view (if selected) when the pad-
lock icon is unlocked. Press the Zero key to enter a zero pressure offset.
6.5.2.3.2 Head Correction
Head correction can be adjusted on this tab when the padlock icon is unlocked. Press
the Head Correction key to modify the head pressure correction settings.
Four parameters may be set to reflect conditions at the operator’s site. Press either
English or Metric units for entering head pressure correction information into the CPG
2500. Press the Head Correction key and the following screen will appear: