Mensor/WIKA Operating Instructions Series CPG 2500
41
Digital Pressure Gauge
SERIES CPG 2500
DOC
mm/dd/yyyy
Sets the date of cal for the active channel.
DOC?
<sp>mmddyy<cr><lf>
Returns the date of cal for the active chan-
nel.
DOM?
<sp>mm/dd/yyyy<cr><lf>
Returns the date of manufacture.
FILTER
Off, Low, Normal, High
Sets the reading filter 0, 80%, 92%, 95%.
FILTER?
<sp> (filter)<cr><lf>
Returns the reading filter.
GASDENSITY
Value in lb/cuft
Sets the head pressure gas density.
GASDENSITY?
<sp>n.nn<cr><lf>
Gets the head pressure gas density.
GASTEMP
Value in degrees F
Sets the head pressure gas temperature.
GASTEMP?
<sp>n.nn<cr><lf>
Gets the head pressure gas temperature.
GATEWAY
nnn.nnn.nnn.nnn
Sets the Ethernet gateway address.
GATEWAY?
<sp>nnn.nnn.nnn.nnn
<cr><lf>
Gets the Ethernet gateway address.
HEIGHT
Value in inches
Sets the head pressure height.
HEIGHT?
<sp>n.nn<cr><lf>
Gets the head pressure height.
ID?
<sp> MENSOR, 2500,
ssssss,v.v.vv
Ssssss is the serial number, v.v.vv is the
CPG 2500 software version.
IP
nnn.nnn.nnn.nnn
Sets the IP address of the instrument.
IP?
<sp>nnn.nnn.nnn.nnn
<cr><lf>
Returns the IP address of the instrument.
KEYLOCK
yes or no
Locks or unlocks keyboard.
KEYLOCK?
<sp>(yes or no)<cr><lf>
Returns yes or no.
LISTRANGE?
PRI,1, min,max
Returns the range of the installed sensor
for the active channel.
LOCALGRAVITY
value in ft/s^2
Sets the local gravity in feet/sec^2.
LOCALGRAVITY?
<sp>n.nn<cr><lf>
Returns the local gravity in feet/sec^2.
NETMASK
nn.nnn.nnn.nnn
Sets the Ethernet network mask.
NETMASK?
<sp>nnn.nnn.nnn.nnn
<cr><lf>
Gets the Ethernet network mask.
OUTFORM
1 to 4 - (see Table 7.1.3.6)
Sets the output format.
OUTFORM?
<sp>X<cr><lf>
Returns the output format (see 7.1.3.4).