Mensor/WIKA Operating Instructions Series CPG 2500
55
Digital Pressure Gauge
SERIES CPG 2500
The ‘Q’ Command Data/Output Format
n
Description
Output Format
0
Pressure reading in the se-
lected units. This is the default
output format.
NNNNNNN<cr><lf>where each N is a number (0
through 9), +,-,decimal point, or a space. If the pres-
sure rate or peak monitor option is being used the
output format is NNNNNNN,NNNNNNN<cr><lf>. The
data before the comma is the pressure and the data
after the comma is the pressure rate or peak.
1
Raw A/D readings (for factory
use)
nnnnnnn,nnnnnnn<cr><lf> where each n is a number
(0 through 9) or a space.
2
Unit ID
Unit ID for the CPG 2500
3
Pressure range and measure-
ment units
NNNNNNN,NNNNNNN, @@@@@@@<cr><lf> where
each N is a number (0 through 9), decimal point, or a
space and each @ is an alpha-numeric character. The
output string represents the minimum pressure, maxi-
mum pressure and the pressure units.
4
Error status code
CPG 2500’s error codes
5
Calibration data
not emulated
6
Pressure rate or peak indica-
tion
nnnn where each n is a number (0 through 9) valid only
if optional pressure rate or peak is enabled. The mean-
ing of each digit is shown under the related feature in
the Options section of the DPG II manual.
7
Zero and span corrections
ZZZZZZZ,SSSSSSS<cr><lf> where ZZZZZZZ is the
zero correction and SSSSSSS is the span correction.
8
Calibration coefficients
not emulated
S Command:
same as SPAN command.
T Command:
used with RATE option; similar to DISPLAY command.