32
Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
i
Notice
Head correction parameters are stored separately for each channel.
Height:
Enter the difference in height between the center of the measure/control port of
the CPG 2500 and the reference level of the Device Under Test (DUT). If the reference
level of the DUT is lower than the center of the measure/control port of the CPG 2500,
enter a positive height. If it is higher, enter a negative height.
Gas Density:
If nitrogen (N2) or dry air are being used as a pressure media, press the
appropriate selection. If another gas is being used, enter the density for the gas at stan-
dard pressure and temperature in either lb/cubic foot (english) or kg/liter (metric) units.
Gas Temperature:
Enter the average gas temperature in degrees F or C. If unsure of
the gas temperature use 68 F.
Local Gravity:
Enter the local gravity acceleration value. If unsure, press the Standard
gravity key.
Limits:
Height
±1200 inches
Density
0 to 1 lb/cu ft
Temperature
0 to 120 °F
Gravity
32 to 32.4 ft/sec²
6.5.2.4 Data Tab
Press the data tab to display the sensor data.