Mensor/WIKA Operating Instructions Series CPG 2500
77
Digital Pressure Gauge
SERIES CPG 2500
10.5.2.2 Restore Factory Cal
Press OK to restore the factory settings.
10.5.2.3 Change Password
1.
Press this key to change the password. A number entry screen will be displayed. En-
ter from one to six digits for a new password. As each number is pressed it appears
in the password window.
!
Caution
CAUTION: The password is seldom used and is easily forgotten.
After a change, write down aned save the new number. If the pass-
word is lost, contact mensor.
2.
Before proceeding, review the displayed digits for accuracy. A mistake here could
prevent future access to this screen.
3.
When satisfied that the new password is correct, and a written copy has been
stored, press Enter to complete the entry.
4.
Confirm the new password by pressing the back key, then returning to the Calibra-
tion screen. If this screen can’t be accessed by using either the old or the new pass-
word contact Mensor.
10.5.2.4 Date of Calibration
Press the ‘Date of Calibration’ to enter the date of calibration in a (mm/dd/yyyy) format.
Press the back key to save the new date of calibration in the sensor.