50
Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
SAVE
SAVElf
Send a SAVE command to preserve changes
made to variable parameters. Without the SAVE
command recent changes will remain in effect only
as long as the power is continuous, or until later
changes displace them. If there is a power inter-
ruption (power OFF) before the SAVE command
is issued, the next power up will return the instru-
ment to its last saved settings.
SEA_LEVEL
SEA_LEVEL sp
<value>lf
SEA_LEVEL sp
<text>lf
[Baro]: This command is used to insert a local
elevation correction for barometric pressure read-
ings. The command must be preceded by either
the PWSL or the PW password. If the adjustment
command is sent before the required password,
“SEA LEV CAL DISABLED” will appear briefly
on the display and the adjustment value will not
be accepted. The command value is in FEET if
pressure units are English or a value in METERS if
pressure units are Metric. The sea level correction
may not be immediately reflected in the output. A
SEA_LEVEL switch command allows the correc-
tion to be alternately applied to, or removed from
the pressure reading.
Send SEA_LEVEL sp <text> lf where <text> is:
y
to have the correction added to the output;
n
to remove the correction from the output, but
still retain the latest value for future use.
Use the SAVE command to save both the sea level
value and the switch settings.
SEA_LEVEL?
SEA_LEVEL? lf
[Baro]: Returns the current elevation correction.
Returns: <value><text><cr><lf> where <value> is
the sea level correction value and where <text> is
FEET or METERS (units).