Mensor/WIKA Operating Instructions Series CPG 2500
43
Digital Pressure Gauge
SERIES CPG 2500
RWINDOW?
<sp>n.nn<cr><lf>
Returns rate exponential filter window.
SBAUD
9600, 19200, 38400, 57600
Sets the serial baud rate.
SBAUD?
<sp>XXXX<cr><lf>
Returns the serial baud data.
SCREENSRC
A, B, or A,B, or B,A, or
Delta,A, or Delta,B
Sets the CPG 2500 display mode.
SCREENSRC?
<sp>{A,B or A,Delta or
Delta,B}<cr><lf>
Returns screen source as one of the
above options.
SDATA
7 or 8
Sets the serial data bits.
SDATA?
<sp>X<cr><lf>
Returns the serial data bits number.
SENSORID?
<sp>Mensor QRS,SN
XXXXXX,VER V.VV
Returns the active sensor’s serial number
and firmware version.
SPAN
desired pressure or ?
Sets span on active channel or for ?,
clears previous value, must be >50% FS
and has a 1% limit.
SPAN?
<sp>XXXXXXX<cr><lf>
Returns span scale factor for active chan-
nel.
SPARITY
even, odd, none
Sets the serial parity.
SPARITY?
<sp>CCCC<cr><lf>
Returns the serial parity.
SSTOP
1 or 2
Sets the serial stop bits.
SSTOP?
<sp>X<cr><lf>
Returns the serial stop bits.
UNITS
units code or text in Table
7.1.3.6.
Sets the engineering units for the active
channel.
UNITS?
<sp>CCCC<cr><lf>
Returns the pressure units of the active
channel in a text string.
UNLOCK
none
Releases Acquire locks. See ACQUIRE?
and RELEASE?
WINDOW
value in current units
Sets the exponential filter window for the
active channel.
WINDOW?
<sp>n.nn<cr><lf>
Returns the exponential filter window for
the active channel.
ZERO
desired pressure or ?
Sets zero to set pressure or for ?, clears
previous value.
ZERO?
<sp>xxxxxxx<cr><lf>
Returns zero offset for active channel.
Summary of Contents for SERIES CPG 2500
Page 87: ......