52
Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
UNITS_TABLE
not emulated
UNITS_TABLE?
UNITS_TABLE? lf
not emulated
WINDOW
window sp
<value>lf
Sets the filter window for pressure readings where
<value> is a value in the current measurement
units within the range of the instrument.
WINDOW?
window?lf
Returns the filter window setting for the pressure
reading.
The instrument returns <value><cr><lf> where
<value> is a value in the current units within the
range of the instrument.
ZERO
zero sp <value>lf
This command requires that the PWZ or the PW
password must be issued before the first occur-
rence of the zero command. The zero command
sets the DPG zero offset to <value> in the current
units where <value> is a value b/- 17 psi.
ZERO?
ZERO?lf
Returns the current zero calibration offset where
<value> is current zero value in current measure-
ment units.
7.1.3.8.2 UNITS Command Syntax for DPG 2100 Measurement Units
Table 7.1.3.8.2 - UNITS Command Syntax for DPG 2100 Measurement Units
N
Description
Output Format
Type
1
pounds per square inch
PSI
English
2
inches of mercury @ 0°C
INHG
English
3
inches of mercury @ 60°F
INHG
English
4
inches of water @ 4°C
INH2O
English
5
inches of water @ 20°C
INH2O
English
6
inches of water @ 60°F
INH2O
English
7
feet of water @ 4°C
FTH2O
English
8
feet of water @ 20°C
FTH2O
English