5.4 Rear Panel
Up to five pneumatic pressure ports are located on the rear panel. Positioned between
the pressure ports are the ethernet and RS-232 connectors and the power switch. Op-
tional connectors could be a GPIB connector, a barometric reference hose barb, or an
analog output connector.
5.5 Pressure Connections
i
Notice
When making up a connection to an o-ring adapter fitting port use a
back-up wrench to prevent over-stressing the threads in the mani-
fold block.
All of the primary pressure ports on the rear are female 7/16 - 20 SAE/MS straight
threads per MS16142 and SAE J514 table 14. They require a tube fitting boss seal with
an o-ring per MS33656. Mensor provides female 1/8” NPT adapter fittings with the
instrument. The pressure connections can be made to these adapters with the proper
mating hardware. We recommend the use of either Loctite® Hydraulic Sealant or new
teflon tape on the threads of any male pipe fitting. Do not use sealant on fittings sealed
with an o-ring. The integrity of each seal is particularly important since even microscop-
ic leaks can cause errors in pressure measurements.