CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Page 9.23
9.4.1.3.
Default Setting
To load the default parameter values, scroll using the
+/-
effort level keys until the display shows:
DEFAULT SETTING ?
press
ENTER
to write the default values to the low kit, and return to the upper menu level by
pressing
CLEAR
for a few seconds.
To write these parameters to the low kit, use the “Write to low kit” function.
9.4.1.4.
Table of configuration parameters
Parameter Description
Defaultvalues
Par 01
Transmission ratio
4248
Par 02
Speed ratio (integer part)
400
Par 03
Minimum braking current (standby pos.) [mA]
500
Par 04
Maximum continuous power [W]
200
Par 05
Machine torque correction coefficient (constant torque)
85
Par 06
User friction torque
10
Par 07
Maximum coil temperature [°C]
125
Par 08
Minimum revolutions [RPM]
35
Par 09
RPM integration scale
400
Par 10
Filter threshold [RPM]
400
Par 11
Speed ratio (decimal part)
0
Par 12
User friction torque delta
30
Par 13
NOT USED
48