CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Pagina 7.23
Figure 7.11-3
5.
Unscrew the tie-rod
(c)
from the fastener on
the belt using an 8mm wrench, without
moving the nut
(d)
.
6.
Separate the tie-rod from the fastener on the
belt, pulling it out as illustrated in the figure
During reassembly, lock down the tie-
rod (c) using a torque wrench setting of
3.7Nm/2.72ft.lbf, until the nut is fully up
against the clamping system. Lock down
the nut using a torque wrench setting of
18Nm/13.3ft.lbf.
If the nut has been moved from its
original position, check in paragraph:
8.6. “Tie rod length”.
Figure 7.11-4
7.
Disassemble the tie-rod fastener from the
belt, by sliding component
(e)
along the belt.
8.
Then remove the wedge
(f)
and the threaded
connector
(g)
from the belt.
Figure 7.11-5
To reassemble the belt, carry out the above steps
in reverse order.
The cams used on the two sides of the
machine, are both the same part so
check the proper point of connection for
spring and belt.
Summary of Contents for CLASS & TREND