Routine Maintenance
81
5
•
Small, flat-head screwdriver
•
Small, Phillips-head screwdriver
•
Clean, 1000-mL beaker
•
Clean, 500-mL graduated cylinder
Note:
Use only glassware you have not cleaned with surfactants.
•
HPLC-grade methanol
•
HPLC-grade water
•
Formic acid
•
Ultrasonic bath
•
Source of oil-free, inert gas (nitrogen or helium) for drying (air-drying optional)
•
Lint-free paper towels
Spare Parts
You might need to replace these spare parts when cleaning source components.
Source Cleaning Procedure
1.
Disassemble the source components (see
“Disassembling Source Components” on
2.
Clean the source components (see
“Cleaning the Source Components” on page
).
3.
Remove and clean the hexapole assembly (see
4.
Reassemble the source components (see
“Reassembling the Source Components”
Item
Waters Part Number
Reference Number
Ion block D-ring (AS035)
700001192
5711312
Viton O-ring (AS214)
700000902
5711294
Extraction cone O-ring
700001897
5711967
Sample cone O-ring
700001194
5711321
Summary of Contents for Micromass ZQ 2000