Installing
39
2
Figure 2-8 ESI Probe in Situ
2.3.6 Connecting the Workstation
Waters ships the workstation with preinstalled MassLynx software. Before connecting the
workstation to the instrument, set it up according to its accompanying instructions. You
should locate the workstation within 16 feet (5 meters) of the instrument.
1.
Connect one end of the network cable to the appropriate port on the rear panel of
the detector.
2.
Connect the other end of the network cable to the port labeled ZQ on the
workstation rear panel.
Caution:
Do not connect the instrument’s power supply cord until you complete the
installation procedures in the previous sections.
Probe Cable
Desolvation Gas Port
Source Cable
Nebulizer Gas Port
Knurled Thumbscrews (2)
Probe
Source Cover
Source Cover Clip
Flange
Probe Adjustor Plate
Probe Adjustor
Source Cover Clip