Setting Calibration Parameters
56
4
3.
Enter these suggested initial reference solution values in Tune window’s Peak
Editor.
Note:
These settings are offered as reference points only and, once adopted, might
require adjusting.
4.
Select rows 1 to 4 in the Peak Editor. This specifies four mass peaks in the Tune
window’s Peak Display area (see
).
5.
Click the
ES+ Source
tab, and enter these suggested parameters in the
corresponding fields of the ES+ Source page (
Row
Mass (ZQ 2000)
Mass (ZQ 4000)
Span
Gain
1
172.9
172.9
5
8
2
772.5
1521.9
5
20
3
1521.9
2271.4
5
40
4
1971.6
3470.5
5
39
Parameter
Suggested Value
Capillary (kV)
3.5
Cone (V)
50
Extractor (V)
3
RF Lens (V)
0.5
Source Temperature (
o
C)
80
Desolvation Gas Flow Rate (L/hr)
250
Desolvation Temperature (
o
C)
120
Cone Gas Flow (L/hr)
50
Caution:
Failure to flow desolvation gas during ESI operation can damage
the source.