Installing the Detector
38
2
Figure 2-7 Rheodyne Injector
2.3.5 Installing the ESI Probe
Refer to
for this procedure.
1.
Connect the PTFE tubing from the probe adjustment flange to the desolvation gas
port on the front panel.
2.
Remove the protective sleeve, if fitted, from the electrospray probe tip.
3.
Slide the probe into the hole in the probe adjustor plate until the probe body rests on
the probe adjustment flange. The probe identification contacts must touch the
screws on the probe adjustment flange.
4.
Secure the probe with the two knurled thumbscrews.
5.
Connect the 4-mm PTFE tubing from the probe to the nebulizer gas port.
6.
Connect the electrical lead from the probe to the capillary connector on the front
panel.
5
6
1
2
3
4
(To Source)