Calibrating
65
4
Figure 4-10 Calibration Acquisition Setup Dialog Box
Table 4-1 Recommended Calibration Acquisitions Setup Parameters
Parameter
ZQ 2000
ZQ 4000
Scan From (amu)
50
50
Scan To (amu)
2040
4080
Run Duration (mins)
1.0
2.0
Data Type
Continuum
Continuum
Static Span ± (amu)
4
4
Static Dwell (sec)
0.1
0.1
Slow Scan Time (sec)
10
20
Fast Scan Time (sec)
0.40
0.90
Inter Scan Delay (sec)
0.1
0.1