Calibrating
55
4
3.
Select
File > Open
. The Open dialog box appears (
).
Figure 4-2 Open Dialog Box
4.
Select
Uncal.cal
, and click
Open
. The Calibration window reappears.
5.
Make sure the phrase “No calibration” follows the three calibration types: Static,
Scanning, and Scan Speed Compensation.
4.2 Setting Calibration Parameters
The rest of this chapter describes how to specify calibration parameters in these windows
and dialog boxes:
•
Tune window
•
Instrument Threshold Settings dialog box
•
Automatic Calibration Check dialog box
•
Calibration Parameters dialog box
•
Mass Measure dialog box
•
Automatic Calibration dialog box
•
Calibration Acquisition Setup dialog box
4.2.1 Tune Window Settings
1.
Follow the syringe preparation procedure in
, this time loading the
syringe with sodium cesium iodide solution (from the API Test Kit).
2.
Click
(Syringe pump) in the Tune window to infuse the solution into the source.