Installing
31
2
2.1 Site Selection and Power Requirements
Install the detector on a stable, level, and appropriately clean surface that meets the
specifications in
.
Table 2-1 Installation Site Requirements
Factor
Requirement
Temperature range
15 to 28
°
C (59 to 82.4
°
F)
Relative humidity
range
20 to 80%, noncondensing
Bench space
Width: 15.2 in. (38.1 cm)
Depth: 26 in. (66.1 cm)
Height: 23 in. (57.2 cm)
Weight: 210 lb. (95.3 kg)
Clearance
Rear: 4.75 in. (120 mm)
Right side: 20 in. (0.5 m) to allow for service access.
Note:
Movable equipment can be located as close as 4.75 in.
(120 mm).
Left side: 0.0 in./mm
Top: 11 in. (28 cm)
Power requirements
Grounded AC, 230 V, 50/60 Hz
Electromagnetic fields No nearby source of electromagnetic noise, such as NMR
systems or magnetic sector mass spectrometers
Static electricity
Negligible
Vibration
Negligible
STOP
Attention:
To avoid overheating the instrument, allow the prescribed clearances at the
top, sides, and rear of the instrument.
STOP
Attention:
To avoid damaging the instrument, measure the voltage at the 230 VAC outlet
before installing the instrument. Readings of less than 208 VAC indicate the need for a
step-up transformer.