Using the ESCi Multi-Mode Ionization Source
113
A
3.
Optimize the fast-transitioning values,
ES Voltages
,
APCI Voltages
, and
Source
Voltages
, for each mode and polarity (
4.
Optimize the slow-transitioning values,
Temperatures
and
Gas Flow
, for both ESI
and APCI performance (
).
Note:
The ESCi+ and ESCi– pages should display identical
s
low-transitioning
values.
5.
Select
APcI
or
ESI
for each row in the Ion Mode column by clicking the drop-down
list in each row (
Note:
You cannot select polarity from the Ion Mode column. Instead, click
Ion Mode
in the menu bar, and select mode and polarity from the drop-down list. This
introduces the ESI+, ESI–, APCI+, or APCI– Source page, which determines an
acquisition’s polarity.
Figure A-7 Selecting the Ion Mode in the MassLynx Peak Editor
shows the Tune window as it appears while the instrument operates in ESCi
mode.
Note:
In the peak display, the APCI peaks appear in blue, the ESI peaks in red
(
).
Peak
Fast-
Display
Transitioning
Variables
Slow-
Transitioning
Variables
Ion Mode
List