Maintaining
82
5
Disassembling Source Components
The source components comprise the pumping block and ion block assemblies, which you
must disassemble before cleaning. Consult
and
when disassembling
and assembling these assemblies.
1.
Stop the liquid flow, then disconnect the LC line from the probe.
Figure 5-5 Pumping Block Assembly
STOP
Attention:
Never remove either probe before the source heater cools to
below 100 °C and the source heaters cool to below 50 °C. Doing so will
shorten the probe heater life. To reduce cooling time, continue flowing API
gas.
TP02001
1
2
3
4
5
6
7
8
10
11
12
1 – Ion Block Assembly
2 – O-Ring, Viton (AS016)
3 – O-Ring, Viton (2-030)
4 – Ion Block Support
5 – O-Ring, Viton (2-039)
6 – O-Ring, Viton (AS016)
7 – O-Ring, Viton 42.0 × 1.5 mm
8 – PTFE Encapsulated O-Ring
9 – Source Pogo Pins
9
10 – PEEK Connector Body
11 – O-Ring, Viton 2-103
12 – Threaded Plug
13
13 – Pumping Block
14 – Hexapole
14