4 Maintenance Procedures
64
December 6, 2012, 71500126802IVD Rev. C
To empty the exhaust trap bottle
1. In the ACQUITY UPLC Console, click Stop Flow
to stop the LC
flow.
2. In the ACQUITY UPLC Console, click API
to stop the desolvation
gas flow.
3. Unscrew and remove the exhaust trap bottle cap.
4. Dispose of the waste liquid in accordance with local environmental
regulations.
5. Fit and tighten the exhaust trap bottle cap.
6. Ensure that the exhaust trap bottle is secured in the upright position.
7. In the ACQUITY UPLC Console, click API
to start the desolvation
gas flow.
8. In the ACQUITY UPLC Console, click Start Flow
to start the LC
flow.
Warning:
The waste liquid in the exhaust trap bottle
comprises ACQUITY UPLC solvents and analytes. Always
wear chemical-resistant, powder-free gloves while handling
the exhaust trap bottle.
Warning:
The waste liquid can be contaminated with
biohazardous and/or toxic materials. Ensure that it is
correctly disposed of according to local environmental
regulations.
Warning:
To confirm the integrity of the source exhaust
system, the following leak test must be performed.
Caution:
To avoid damage to the instrument, snoop (or
equivalent) leak detector liquid must be used only for the
purpose described in the following step; it must not be used on
any other part of the instrument.
Summary of Contents for TQ
Page 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...
Page 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...
Page 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...
Page 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...
Page 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...