Connecting the oil-filled roughing pump
December 6, 2012, 71500126802IVD Rev. C
165
3. Connect the relay box power connector to the main power source.
Making the electrical connections for a roughing pump without an
external relay box
Roughing pump connections without an external relay box
To make the electrical connections for a roughing pump without an external
relay box
1. Connect the roughing pump power cord to the main power source.
2. Connect the relay cable from the roughing pump d.c. connector to the
pump connector on the instrument’s rear panel.
01757 U.S.A.
!
!
V ~ 200 - 240
V ~ 200 - 240
Hz
50 - 60
Hz
50 - 60
V
VA
900
ACN 065444751
1
2
3
4
5
6
7
8
9
10
A nalog
N ot used
Stop Flow
Sw itch 2
G round
G round
O ut
O ut
1
2
3
4
5
6
7
8
9
10
Inject S tart
Event
Sw itch 3
Sw itch 4
G round
G round
In
In
O ut
O ut
O ut
RS 232
RS 232
ETHERNET
ETHERNET
API Gas
API Gas
COLLISION GAS
COLLISION GAS
6.9 Bar Maximum
6.9 Bar Maximum
1.0 BAR MAXIMUM
1.0 BAR MAXIMUM
SOURCE VENT
SOURCE VENT
VACUUM
VACUUM
VACUUM
VACUUM
IVD
IVD
PUMP
PUMP
Serial Number
RP00004
Roughing pump main
power connector
To power source
Detector rear panel
Roughing pump d.c. connector
Summary of Contents for TQ
Page 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...
Page 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...
Page 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...
Page 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...
Page 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...