4 Maintenance Procedures
128
December 6, 2012, 71500126802IVD Rev. C
•
Source enclosure housing seal
•
Source enclosure side flange seal
•
Probe adjuster assembly flange seal
9. Fit the new probe adjuster assembly probe seal to the probe adjuster
assembly.
Assembling the probe adjuster assembly and source enclosure
Required materials:
•
Chemical-resistant, powder-free gloves
•
4-mm Allen wrench
To assemble the probe adjuster assembly and source enclosure
1. Fit the source enclosure side flange to the source enclosure.
2. Fit and tighten the 3 thumbscrews that secure the source enclosure side
flange to the source enclosure.
3. Fit the probe adjuster assembly to the source enclosure.
4. Use the 4-mm Allen wrench to fit and tighten the 4 screws securing the
probe adjuster assembly to the source enclosure.
Caution:
The source enclosure side flange securing thumbscrews
must each be sequentially and incrementally tightened until
they are all fully tight.
Caution:
The probe adjuster assembly securing screws must each
be sequentially and incrementally tightened until they are all
fully tight.