4 Maintenance Procedures
132
December 6, 2012, 71500126802IVD Rev. C
Air filter inside the instrument’s lower bezel
3. Reinstall the lower bezel.
Replacing the air filter inside the lower bezel
If you cannot clean the air filter by vacuuming, replace it with a new filter.
Required materials:
•
T10 TORX driver
•
TQ Detector lower bezel air filter
To replace the air filter inside the lower bezel
1. Remove the 5 screws that secure the lower bezel in place (see the figure
2. Use the T10 TORX driver to remove the 4 screws that secure the air
filter frame and air filter to the inside of the lower bezel.
TP02638
Air filter
Frame
Summary of Contents for TQ
Page 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...
Page 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...
Page 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...
Page 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...
Page 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...