Maintaining the instrument air filters
December 6, 2012, 71500126802IVD Rev. C
131
Maintaining the instrument air filters
Cleaning the air filter inside the instrument’s lower bezel
Required materials:
•
T10 TORX driver
•
Vacuum cleaner
To clean the air filter inside the instrument’s lower bezel
1. Use the T10 TORX driver to remove the 5 screws that secure the lower
bezel in place.
Instrument’s lower bezel
2. Use a vacuum cleaner to clean the air filter located inside the
instrument’s lower bezel.
TP02636
Screws
Screws