B External Connections
184
December 6, 2012, 71500126802IVD Rev. C
Signal connections
Requirement:
To meet the regulatory requirements of immunity from
external electrical disturbances, you must install connection covers over the
signal connectors.
To make signal connections
1. Reference the signal connection location from the silk-screened label for
inject start or any other input/output connection you plan to use from
Connector I or II on the rear panel of each instrument.
Instrument analog-out/event-in connections
Signal connections
Description
Analog (Out)
Used for analog chart output functionality. The
output voltage range is 0 to 2 V. The resolution of
the voltage output is 12 bits.
Stop Flow (Out)
Used to stop the solvent flow if the nitrogen gas
supply fails. Maximum 30 V, 0.5 A, 10 W.
Inject Start (In)
Signals the start of an injection. Maximum 30 V.
Event (In)
Allows an external device to start data acquisition.
Maximum 30 V.
Switch 2 (Out)
Used to send time-based contact closure signals to
external devices. Maximum 30 V, 0.5 A, 10 W.
Switch 3 (Out)
Used to send time-based contact closure signals to
external devices. Maximum 30 V, 0.5 A, 10 W.
Switch 4 (Out)
Used to send time-based contact closure signals to
external devices. Maximum 30 V, 0.5 A, 10 W.
Summary of Contents for TQ
Page 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...
Page 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...
Page 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...
Page 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...
Page 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...