4 Maintenance Procedures
124
December 6, 2012, 71500126802IVD Rev. C
Removing the seals from the source enclosure and probe
adjuster assembly
Required materials:
•
Chemical-resistant, powder-free gloves
•
Flat-blade screwdriver
•
O-ring removal kit
To remove the seals from the source enclosure and probe adjuster
assembly
1. Use the O-ring removal kit to carefully remove the probe adjuster
assembly probe seal from the probe adjuster assembly (see
).
2. Use the O-ring removal kit to carefully remove the following seals from
the source enclosure:
•
Source enclosure housing seal
•
Source enclosure side flange seal
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
TP02632
Probe adjuster assembly probe seal