Replacing the source assembly seals
December 6, 2012, 71500126802IVD Rev. C
125
•
Probe adjuster assembly flange seal
3. Open the source enclosure door by releasing both spring-clips and
lowering the door towards you.
Caution:
Do not apply any downward force to the source
enclosure door while the door is open.
RP00025
Source enclosure
housing seal
Source side
flange seal
Probe adjuster
assembly flange seal