Replacing the source assembly seals
December 6, 2012, 71500126802IVD Rev. C
121
Replacing the source assembly seals
To avoid possible excessive leakage of solvent vapor into the laboratory
atmosphere, the following seals must be renewed at intervals of no greater
than 1 year:
•
Source enclosure door seal
•
Source enclosure door glass seal
•
Source enclosure housing seal
•
Source enclosure side flange seal
•
Probe adjuster assembly probe seal
•
Probe adjuster assembly flange seal
To complete this procedure, you must pressure test the source, as described in
the
Waters Micromass Source Pressure Test Unit Operator’s Guide
.
Removing the source enclosure from the instrument
Required materials:
•
Chemical-resistant, powder-free gloves
•
5-mm Allen wrench
To remove the source enclosure
1. Vent and shut-down the instrument (see the mass spectrometer’s online
Help for details)
.
Warning:
To avoid possible excessive leakage of solvent vapor
into the laboratory atmosphere, the seals listed below must be
renewed, at intervals of no greater than 1 year, exactly as
described in this section.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The source can be hot. To avoid burn injuries, allow it
to cool for at least 30 minutes before proceeding.
Summary of Contents for TQ
Page 34: ...1 Waters TQ Detector 34 December 6 2012 71500126802IVD Rev C...
Page 144: ...4 Maintenance Procedures 144 December 6 2012 71500126802IVD Rev C...
Page 156: ...A Safety Advisories 156 December 6 2012 71500126802IVD Rev C...
Page 188: ...B External Connections 188 December 6 2012 71500126802IVD Rev C...
Page 192: ...C Materials of Construction and Compliant Solvents 192 December 6 2012 71500126802IVD Rev C...