Cotter Corporation (N.S.L.)
Cañon City Milling Facility
Cañon City, Colorado
Procedure
Oxidation Reduction
Potential Measurements
Number: E-046
Page 6 of 6
Revision: 1
Date: --/--/--
Appendix 5.1
ORP Reference Solution Offset Correction Addition Factors
Temp
(
0
C)
ORP
Standard
Solution
*(mV)
Eh (SHE)
(mV)
Correction
Addition
Factor
Temp
(
0
C)
ORP
Standard
Solution
(mV)
Eh (SHE)
(mV)
Correction
Addition
Factor
Temp
(
0
C)
ORP
Standard
Solution
(mV)
Eh (SHE)
(mV)
Correction
Addition
Factor
0
218
438
220
17
219
426
207
34
220
412
192
1
218
437
219
18
219
425
206
35
220
411
191
2
218
437
219
19
219
424
205
36
220
410
190
3
218
436
218
20
219
424
205
37
220
409
189
4
218
435
217
21
219
423
204
38
220
408
188
5
218
435
217
22
219
422
203
39
220
407
187
6
218
434
216
23
219
421
202
40
220
406
186
7
218
433
215
24
220
420
200
41
220
405
185
8
218
433
215
25
220
420
200
42
220
404
184
9
219
432
213
26
220
419
199
43
220
403
183
10
219
431
212
27
220
418
198
44
220
402
182
11
219
430
211
28
220
417
197
45
220
401
181
12
219
430
211
29
220
416
196
46
220
400
180
13
219
429
210
30
220
415
195
47
220
399
179
14
219
428
209
31
220
414
194
48
220
398
178
15
219
428
209
32
220
413
193
49
220
397
177
16
219
427
208
33
220
412
192
50
220
396
176
Note: * Using pH/mV electrode with 4M KCL Ag/Ag CL reference solution