13
sympHony
™
Meter User Guide
Top Line
Middle Line
Bottom Line
Setup Menu Description
(default setting, method specific)
rEAd
tyPE
AUt0, tImE, C0nt,
Measurement read type as AUTO-READ, timed or continuous
(AUt0, yes)
rEAd
tInE
00:05 to 99:59
Timed measurement value in minutes and seconds
(01:00, yes)
L0g
dEL
n0, YES
Delete datalog after download option
(n0, yes)
gEn
dEgC
-5.0 to 105
Manual temperature value
(25.0, yes)
gEn
StIr
0FF, 1, 2, 3, 4, 5, 6, 7
Stirrer speed – benchtop meters only
(4, yes)
gEn
AUt0
0n, 0FF
Automatic meter shutoff option
(0n, no)
gLP
SEt
0FF, 0n
GLP option, GLP feature enables or disables methods
(0FF, no)
gLP
PASS
0000 to 9999
Password entry
(0000, no)
dAtE
H0Ur
HH00 to HH23
Hour setting
(HH12, no)
dAtE
tInE
mm00 to mm59
Minute setting
(mm00, no)
dAtE
tYPE
mdY, dmY
Displayed date format as month, day, year or day, month, year
(mdY, no)
dAtE
YEAr
2000 to 2099
Year setting
(2004, no)
dAtE
dAtE
mm01 to mm12
Month setting
(mm01, no)
dAtE
dAY
dd01 to dd31
Day of the month setting
(dd01, no)
r232
bAUd
1200, 2400, 4800, 9600
Baud rate setting
(9600, no)
r232
0UtF
Prnt, C0mP
Output format for printer or computer, C0mP format is comma
delimited
(Prnt, no)
Enabling or Disabling the Setup Menu Password
1. In the measurement mode, press
.
2. Press
/
until
is displayed on top line.
3. Press
to move the arrow icon to the middle line and press
/
until
is displayed.
4. Press
to move the arrow icon to the bottom line and enter a password by
pressing
/
to enter each digit and
to move to the next digit. Enter
to disable the password and allow unlimited access to the setup menu.
5. Press
to move the arrow icon to the top line.
6. Press
to exit the setup menu and return to the measurement mode.
Meter Setup