Required Equipment
·
(5) BNC (coaxial) cables
·
(4) BNC Tees connectors to connect the BNC cables together
·
Differential Mode Tee
·
Agilent 34401A Voltmeter with valid Calibration Certificate. The Voltmeter is our
sole reference source.
Figure 1
Figure 2
Measurement Details
The calibrated reference is a Agilent 34401A DMM. The VR8500 Drive output is used to
generate a single ended sinusoidal output. The VR8500 Drive output is paired with the
VR8500 Aux output to produce the differential sinusoidal output. The reference output is
nominally 1 volt p-p (0.707VRMS). Each input can be verified both single ended and
differentially. The outputs and the inputs are verified simultaneously.
At frequencies below 5Hz the 34401A is used in 5-1/2 digit mode to acquire the reference at
300 Samples per Second using DC mode. The Maximum minus the Minimum value of the
acquired reference is then multiplied by the square root of 2 for a reference sinusoidal RMS
value.
At frequencies above 5Hz the 34401A is used in 6-1/2 digit mode to directly measure AC
voltage using slow filter setting.
Summary of Contents for VibrationVIEW
Page 15: ...Document Revision Part I ...
Page 17: ...How To Part II ...
Page 96: ...Menu commands Part III ...
Page 142: ...Test Definition Part IV ...
Page 217: ...Control Centers Part V ...
Page 227: ...Accelerometer Toolbars Part VI ...
Page 231: ...Stop Codes Part VII ...
Page 263: ...Typical Views Part VIII ...
Page 268: ...VibrationVIEW Help 255 1995 2005 Vibration Research Corporation All rights reserved ...
Page 269: ...Dialog Boxes Part IX ...
Page 303: ...Windows Elements Part X ...
Page 309: ...APPENDIX Part XI ...
Page 311: ...APPENDIX 298 1995 2005 Vibration Research Corporation All rights reserved ...