This code indicates that the emergency stop button was pressed, causing the controller to shut down.
The emergency stop button is an optional accessory that connects to the remote input connector on
the rear of the
VibrationVIEW
I/O unit.
7.5.23 Input Signal is Clipped
Input Signal is Clipped (
Shock Stop Code
)
This code indicates that the input signal to channel n is exceeding the allowable input voltage range of
+/- 10 volts, and is being distorted by clipping. To solve this problem, use an accelerometer with a
lower mV/G sensitivity rating.
The maximum allowable acceleration readings for various accelerometer sensitivities are shown in the
table below.
Sensitivity
Maximum Acceleration
10 mV/G
1000 G peak
20 mV/G
500 G peak
50 mV/G
200 G peak
100 mV/G
100 G peak
1000 mV/G
10 G peak
7.5.24 Invalid Remote Test
Invalid Remote Test (
Shock Stop Code
)
This code indicates that the "Remote Start/Stop" and "Remote Selection" options are both enabled but
the test number selected using the remote test selection input (Discrete Inputs 4,5,6,7) does not have
a valid test name configured for the current input settings. Select the
menu command to verify that the test names configured there are valid.
See Also:
How to use remote inputs to control VibrationVIEW
How to use remote inputs to monitor your process
7.5.25 Computer Entering Suspend Mode (copy)
Computer Entering Suspend Mode
This code is given when VibrationVIEW is running a test, the computer is running on batteries (or
UPS), and the battery level is low. The test is stopped in order to allow the computer to switch into
suspend or hibernate mode before the batteries are completely exhausted.
7.5.26 Lost Contact
Lost contact with I/O box n (
Shock Stop Code
)
This code indicates that the
VibrationVIEW
I/O box is no longer communicating with the host computer.
Verify that:
·
The Ethernet cables are connected properly.
·
The box is powered.
7.5.27 Max Starting Voltage
Max Starting Voltage (
Shock Stop Code
)
96
40
47
Summary of Contents for VibrationVIEW
Page 15: ...Document Revision Part I ...
Page 17: ...How To Part II ...
Page 96: ...Menu commands Part III ...
Page 142: ...Test Definition Part IV ...
Page 217: ...Control Centers Part V ...
Page 227: ...Accelerometer Toolbars Part VI ...
Page 231: ...Stop Codes Part VII ...
Page 263: ...Typical Views Part VIII ...
Page 268: ...VibrationVIEW Help 255 1995 2005 Vibration Research Corporation All rights reserved ...
Page 269: ...Dialog Boxes Part IX ...
Page 303: ...Windows Elements Part X ...
Page 309: ...APPENDIX Part XI ...
Page 311: ...APPENDIX 298 1995 2005 Vibration Research Corporation All rights reserved ...