OPERATORS MANUAL
ORION II
Orion II 500-10000 Operators
Page 12 of 95
Manual Part no: DOMAON0100 Issue A
April 2008
3.2.6 Purified Water System Polishing Section
70
Ultra Violet Irradiation*
UV70
This can be fitted to the outlet of the CEDI to enhance
bacteria control.
71
Ultrafiltration*
UF71
This can be fitted to the outlet of the CEDI to remove
endotoxins and bacteria from the product water.
72
Microfiltration*
MF72
This can be fitted to the outlet of the CEDI to enhance
bacteria control.
* NB only one of the above three options may be chosen
3.2.7 TOC Monitor
81
TOC Monitor
AE81-01
Anatel TOC 600 to monitor the product water TOC.
82
TOC Monitor
AE82-01
Sievers 500RL to monitor the product water TOC.
83
TOC Monitor
AE83-01
Thornton 5000TOC to monitor the product water TOC.
3.2.8 Outlet Valves
90
Outlet 3-Way Valve
VA90-01
Diverter valve to divert water to the product water tank when
required and recycle to TK30-01 when not. The valve
should be fitted as close as possible to the product water
tank to minimise product water deadleg.
91
Outlet Two 2-Way
Valves
VA91-01 & VA91-02
Two 2-Way valves to divert water to the product water tank
when required and recycle to TK30-01 when not. The
valves should be fitted as close as possible to the product
water tank to minimise product water deadleg.
92
Outlet Three 2-Way
Valves
VA92-01, VA92-02 &
VA92-03
Three 2-Way valves to divert water to two product water
tanks when required and recycle to TK30-01 when not. The
valves should be fitted as close as possible to the product
water tanks to minimise product water deadleg.
3.2.9 Operating Mode
100
Continuous Operation
Unit runs continuously and recycles product water when
Product Water Tank is full.
101
Start / Stop
Unit starts and stops when tank empties and fills. There is,
however, an intermittent recycle of water to prevent long
periods of stagnation.