![Vatech EzScan-i User Manual Download Page 93](http://html1.mh-extra.com/html/vatech/ezscan-i/ezscan-i_user-manual_3616519093.webp)
Page 93
Radiated RF
IEC 61000-4-3
Table 9 in
IEC-60601-
1-2 2014
Table 9 in
IEC-60601-
1-2 2014
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the EzScan
system, including cables,
than the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance:
where P is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter
manufacturer and d is
the recommended
separation distance in
metres (m). Field
strengths from fixed RF
transmitters, as
determined by an
electromagnetic site
survey, should be less
than the compliance
level in each frequency
range. Interference may
occur in the vicinity of
equipment marked with
the following symbol:
Conducted RF
IEC 61000-4-6
3V 0.15-
80MHz
6V in ISM
bands
between
0.15 MHz
and 80 MHz
80% AM at
1KHz
3V 0.15-
80MHz
6V in ISM
bands
between
0.15 MHz
and 80 MHz
80% AM at
1KHz
NOTE: UT is the A.C. mains voltage prior to application of the test level.