User Manual
Electrostatic
discharge (ESD) IEC
61000-4-2
± 8 kV
/Contact
± 2 kV, ± 4
kV, ± 8 kV,
± 15 kV /air
± 8 kV
/Contact
± 2 kV, ± 4
kV, ± 8 kV, ±
15 kV /air
Surface should be wood,
concrete, or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/ burst IEC
61000-4-4
± 2 kV for
power
supply lines
± 1 kV for
input/outpu
t lines
± 2 kV for
power
supply lines
± 1 kV for
input/outpu
t lines
Mains power quality
should be that of a
typical professional
healthcare facility
environment.
Surge IEC 61000-
4-5
± 0.5 kV, ± 1
kV line(s) to
line(s)
± 0.5 kV, ± 1
kV, ± 2 kV
line(s) to
earth
± 0.5 kV, ± 1
kV line(s) to
line(s)
± 0.5 kV, ± 1
kV, ± 2 kV
line(s) to
earth
Mains power quality
should be that of a
typical professional
healthcare facility
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT
(>95% dip in
UT) for
0,5 cycle
40% UT
(60% dip in
UT) for 5
cycles 70 %
UT (30% dip
in UT) for
25 cycles
<5% UT
(>95% dip in
UT) for 5 s
<5% UT
(>95% dip in
UT) for
0,5 cycle
40% UT
(60% dip in
UT) for 5
cycles 70 %
UT (30% dip
in UT) for
25 cycles
<5% UT
(>95% dip in
UT) for 5 s
Mains power quality
should be that
of a typical professional
healthcare facility
environment.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
location in a typical
professional healthcare
facility environment.