16.D/A CONVERTER (OPTION)
56
- Resolution
The D/A converter has a resolution of 1/10000 to 0 to 10V (4 to 20mA).
16-1. Connection of the D/A Converter Options
The terminal block for the D/A converter options is the cage clamp type. Connect according to the
following procedures.
As for wire materials, use parallel 2-core cables, cabtyre cables, or the like.
Connect to the cage clamp type terminal block by using the attached miniscrewdriver.
1) Strip the casing 0.2in (6mm) on the cable to be connected.
Twist the bare wire to fit the terminal hole.
2) Insert the supplied screwdriver into the upper hole and lift
upward.
3) Insert the twisted wires into the lower hole.
4) Pull the screwdriver out from the upper hole.
5) Make sure cable is clamped securely and does not come out
with a slight tug.
Notice
- The D/A converter is optional.
- Do not apply external voltage because it will break.
- Do not short-circuit the voltage output because a failure will occur.
Also, connecting a capacitive load may cause oscillation.
Notice
- Cable can be from 24 to 14AWG (0.2 to 2.5mm
2
).
- It is not necessary to solder the cable wires or to fix a solderless terminal.
- If several cables to be inserted to the same hole, twist those cable wires together and
insert.