4.CALIBRATION
17
4-1. Span Calibration Procedure
Follow the steps below to perform span calibration.
Release the calibration protection.
Set the unit of the load to be displayed.
Determine the decimal point place of the indicated value.
(Can be omitted if there is no change.)
Enter the minimum value of digital increments.
(Can be omitted if there is no change.)
Set the zero point of the loadcell in no-load condition
(with the loadcell unloaded).
Enter the span (gain) point of the loadcell with a load applied to the
loadcell.
Turn on the calibration protection for preventing misoperation.
4-2. Equivalent Input Calibration Procedure
Follow the steps below to perform equivalent input calibration.
Release the calibration protection.
Set the unit of the load to be displayed.
Determine the decimal point place of the indicated value.
(Can be omitted if there is no change.)
Enter the minimum value of digital increments.
(Can be omitted if there is no change.)
Set the zero point of the loadcell in no-load condition
(with the loadcell unloaded).
Register the rated output value and capacity of the loadcell.
Turn on the calibration protection for preventing misoperation.
Lock2 Release
Unit Setting
Zero Calibration
Calibration Protection
Minimum Scale Division
Span Calibration
Decimal Place
Lock2 Release
Zero Calibration
Calibration Protection
Equivalent Input Calibration
Unit Setting
Minimum Scale Division
Decimal Place