11.TIMING CHART
39
11-2. Discharge Weighing (In The Simple Comparison Mode)
The overweight/underweight comparison timing depends on the OVER/UNDER COMP. setting under
COMPARISON. (It is set at “COMPLETE SIGNAL ON” in the illustration.)
The completion signal output timing depends on the COMPLETE SIG. OUT setting under
COMPARISON.
Conditional expression
-
NZ output
ON when the weight value
≦
preset near zero value
-
SP1 output
ON when the net
≧
preset final weight value – preset SP1 value
-
SP2 output
ON when the net
≧
preset final weight value – preset SP2 value
-
SP3 output
ON when the net
≧
preset final weight value – preset CPS value
- UNDER
output
ON when the net < preset final weight value – preset underweight value
- OVER
output
ON when the net > preset final weight value + preset overweight value
Select the weight value for near zero comparison from GROSS or NET under COMPARISON setting
menu.
LO limit
NET
0
Time
t1
t1
t2
t2
t3
t3
NET
GROSS
The near zero signal is used for
discharge completion confirmation
as in the case of feed weighing.
GROSS for high/low limit comparison, and NET for final weight,
before-final-weight,and near zero comparisons.
(In discharge weighing, NET is compared negatively.)
Materials are fed until the HI limit signal is output.
FINAL – SP1
FINAL – SP2
FINAL – CPS
TARE ON input
SP2 output
SP1 output
SP3 output
Comparison
Judging time
LO limit or less
inhibiting time
NZ output
END output
OVER (UNDER)
output
t1: Comparison inhibiting time
COMP.INH.TIME under
COMPARISON
t2: Judging time
JUDGING TIME under
COMPARISON
t3: Completion output time
COMPLETE OUT T. under
COMPARISON