CALIBRATION PROCEDURE
SCXI -1122
This document contains information and step-by-step instructions for
calibrating the National Instruments SCXI-1122 module.
Contents
Equipment and Other Test Requirements ............................................... 2
Test Equipment ................................................................................ 2
Software and Documentation........................................................... 3
Test Conditions ................................................................................ 4
Verifying Analog Input Offsets ................................................ 5
Verifying Excitation Limits...................................................... 9
Adjusting the Module ...................................................................... 10
Adjusting Gain and Offset Errors ............................................. 10
Adjusting Calibration Constants............................................... 13
Adjusting Excitation Limits...................................................... 14
Verifying Adjusted Values .............................................................. 15
™