Maintenance
93
Check
Interval
(recommended)
After commissioning
Check and, if necessary, update the firmware version.
As required
Store the current configuration
As required
Semi-annual check
Check air filter for contamination/clogging.
Every 6 months
Perform calibration (at shorter intervals depending on accuracy re-
quirements)
As required
Check tube connection to the density measuring cell/sensors for
porosity.
Every 6 months
Check membrane on the gas pressure controller for porosity or lack
of elasticity.
Every 6 months
Remove
/clean
thermoelectric battery and heat exchanger.
Every 6 months
Replace the seal on the heat exchanger.
Every 6 months
Check the fan.
Every 6 months
Check the air intake filter (ambient air inlet, fan filter mat).
Every 6 months
Clean
the gas and air nozzles Change O-rings, if necessary.
Every 6 months
Annual check
Replace air filter.
Annually
Replace neoprene tubes inside the combustion calorimeter.
Annually
Clean the complete system.
Annually
Additional information:
Included documents
2
Flow plates are attached on the side of the thermoelectric battery and in the thermal elements of the
combustion calorimeter. The flow plates must not be twisted when removing the thermoelectric bat-
tery. The thermoelectric battery must be pulled out as far as possible and then tilted to the side.
3
Cleaning of the heat exchanger with water followed by thorough drying of the cleaned components.
4
Clean the nozzles using a volatile solvent.
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...