Transport, installation, and acceptance
29
6.3 Installing and connecting
6.3.1 Installation location
The installation location of the combustion calorimeter must meet the following
conditions:
•
Clean room that may be used only for purposes of gas analysis and measure-
ments.
•
At least 50 cm working space on the left side
•
Lockable windows
•
Protected from direct influence by the weather and direct solar radiation
•
Stable room temperature
•
Provided with clean and adequate ambient air for unbiased measurement re-
sult (combustion calorimeter requires approximately 30 m³/h air)
•
Adequate load bearing capacity of the wall is ensured
•
Mounting on a solid wall