Technical data
7
1.5 Display times of combustion value measurement
Dead time:
3 seconds
50% time:
7 seconds
90% time:
CWD2005: 20 seconds,
CWD2005 PLUS: 15 seconds
99% time:
CWD2005: 60 seconds,
CWD2005 PLUS: 55 seconds
1.6 Gas inputs
Process gas:
1, optionally 2 for dual measuring ranges
Calibration inputs:
1, optionally 2 for dual measuring ranges
Test gas inputs:
1
Fast loop:
optional
Gas connections:
Compression fitting 6 mm
1.7 Calibration gas / Test gas
Calibration gas:
Dependent on the process gas
Calibration interval:
Dependent on the process gas
Calibration duration:
Process lasts up to 20 minutes,
gas flow approx. 10 minutes
Gas consumption:
Approx. 7 l gas per calibration (dependent on the
calibration gas)
Input pressure, max:
40 mbar
Input pressure, min.:
20 mbar, dependent on the gas
1
The display times are measured from the time the new gas reaches the burner. Display times are
based on measurements with pure methane.
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...