Operation
78
Range cfg.
is a binary coding for the measurement conditions while a measuring
range switchover is in process.
-1
measuring range not loaded
1
nozzle set 2 or nozzle set 1
2
air nozzle on or off
4
flammable carrier gas on or off
8
not flammable carrier gas (oxygen) on or off
16
flammable carrier gas that does not burn by itself on or off
The number 6 indicates that the measuring range is operated with nozzle set 1 and
a flammable carrier gas with an air nozzle.
The number 2 indicates that nozzle 1 has an air nozzle.
Calibr. gas
indicates the number of the calibration gas for the measuring range.
The upper nozzle set has number 1, the lower nozzle set the number 2.
No.
indicates switchover point 1 and 2.
Signal
indicates the physical unit of the measurement value.
Unit
indicates the switchover point in the specified dimension between the meas-
uring ranges. In the example, the first measuring range is switched over at 650
+5% kcal/m3.
Switch (%)
indicates the overlap of the measuring ranges in %. In the example,
the hysteresis is 5 %.
Measuring range switchover
can be done automatically or manually. If manual
measuring range switchover is set, an additional menu item appears under Num-
ber 6.2.1.3.9. Change range. Here you do the manual switchover to another meas-
uring range.
•
non
•
automatic
•
manual
The time constant of the burner
is a factor determined for every burner. Sewage
gas burners have another factor than burners for natural gas. Every burner type
has its own coefficient.
Max temp of inner pipe
is a switch-off temperature if in case of a fault highly
calorific gas flows to a too large nozzle and generates a significant over-tempera-
ture.
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...