Operation
61
Main menu - Options - I/O - Digital inputs
Various digital signals are configured in this menu.
The following must be observed for this:
The pin assignment is preassigned.
Chapter 6.3.10 Connector assignment Input-Output IOexternal!
Signals
Zero position
Contact
CWD2005
Start calibration
high
open
Calibration starts
closed
No calibration
low
open
No calibration
closed
Calibration starts
Start measurement
high
open
Measuring starts
closed
Measuring stops
low
open
Measuring stops
closed
Measuring starts
Hold signals
high
open
Hold mA
closed
mA online
low
open
mA online
closed
Hold mA
Cancel calibration
high
open
Calibration cancellation
closed
No calibration cancellation
low
open
No calibration cancellation
closed
Calibration cancellation