Transport, installation, and acceptance
33
6.3.4 Process gas
WARNING
Gas connections may only be installed by qualified personnel!
Exhaust gases must be discharged by the owner to a safe environment!
NOTE
•
Connection parts must be clean and free of residues. Contaminants may
reach inside the combustion calorimeter and may cause false measurements
and/or damage.
•
Process gas connection / compression fitting on side of combustion calorim-
eter
•
Gas input pressure depending on gas between 20 - 40 mbar.
Chapter 1 Technical data
•
The input pressure for the gas connections must not exceed the specifica-
tions of the technical data of the combustion calorimeter.
•
Process gas must be free of contaminants and condensate.
•
Each connection point must be carefully checked for leak tightness. When
leaks are present, the system draws air and indicates false measured values.
•
Do not use sealant for sealing the gas connections. Sealant ingredients may
falsify the measurement result.
•
Only suitable lines may be used.
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...