Protective equipment
20
4.3 Thermal cutoff
NOTE
The tripping temperature of the thermal cutoff is 72 °C.
When tripped, the thermal cutoff disconnects the 24 V voltage supply of the sole-
noid valves.
4.4 Solenoid valve
In the event of malfunctions, the solenoid valves of the combustion calorimeter
close automatically. A restart of the combustion calorimeter is necessary.
4.5 Fan
The fan mixes exhaust gases produced during combustion with fresh air and en-
sures an appropriate volume flow. The fan is speed-monitored. If the fan fails, the
combustion calorimeter goes to a safe state.
4.6 Door enclosure
Protective covers of the housing separate electrical components and hot surfaces
of the environment.