Troubleshooting
98
12.4.2 Error/status messages
Filter change
•
Filter change (air filter)
•
The frequency of the frequency converter is > than the set threshold value.
Int. pressure
•
Alarm in the event of inadequate process gas or calibration gas
•
Adjustable threshold; default: 14 mbar
Overtemp
•
The flame is too hot. Wobbe signal > 76 mV,
•
Temperature rise in the inner tube of the thermoelectric battery > 50 °C
Fault
•
Filter change of air filter (the frequency of the frequency converter is > than
the set threshold value)
•
Gas pressure is too low
•
Fault in the thermoelectric battery
•
Defective PT 100 temperature sensor
Service
Filter change of air filter
•
Frequency > threshold value
•
Air inlet temperature > 41 °C
Operation
•
Flame is burning, analyzer is running on process or calibration gas
Operation delay
•
Flame is burning, analyzer is running on process or calibration gas
•
Delay time has elapsed after which the analog signals are enabled
Process
•
Flame is burning, solenoid valve for process gas is open
•
Analyzer is running on process gas
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...